PanOph
Free previewYou're reading the free part of this topic. The rest is included with Clinical.See what's included

Ophthalmic Viscosurgical Devices (OVDs)

Cataract & Refractive Surgery

Teaching images

8

1. Definition, Nomenclature & History

Teaching image

1
Clinical Analogy
  • Think of OVDs as the 'scaffolding' erected inside the eye during surgery - just as a builder uses scaffolding to maintain the structure of a building while work proceeds, the surgeon injects OVD to maintain space, protect delicate surfaces, and stabilize tissues. Once the work is done, the scaffold must be completely removed, or it becomes the source of problems (raised IOP).

Definition

  • Ophthalmic Viscosurgical Devices (OVDs) are sterile, viscoelastic preparations of high-molecular-weight polymers, primarily used intraoperatively to maintain intraocular spaces, protect tissues, and facilitate surgical manoeuvres.
  • The term OVD replaced 'viscoelastic' (VISCOELASTIC) by the FDA in 2000 to emphasize their role as medical devices rather than mere materials.

2. Physical Properties - The Core Science

Teaching image

1
Clinical Analogy
  • Imagine honey vs. water poured into a cup. Honey (cohesive OVD) stays together, pours slowly, and is easy to scoop out entirely. Water (dispersive OVD) spreads everywhere, coats every surface but is difficult to remove completely. This single analogy encapsulates the entire clinical behaviour of OVDs.

Key Rheological Properties

PropertyDefinition & SignificanceClinical Correlate
Viscosity• Resistance to flow at LOW shear rates - determines 'body' and space maintenance• High viscosity → space maintainer | Low viscosity → coater
Pseudoplasticity (Shear thinning)• Decrease in viscosity with increasing shear rate (flow rate) Critical for injectability through small cannulas• All OVDs are pseudoplastic - they thin on injection, thicken at rest
Viscoelasticity• Combination of viscous (liquid) and elastic (solid) behaviour Elastic component allows OVD to reform after deformation• Elastic recovery = space maintenance after phaco turbulence
Zero-shear Viscosity• Viscosity at resting state (no flow) Measured in Pascal-seconds (Pa·s) or mPas• Healon5: ~7,000,000 mPas | Viscoat: ~30,000 mPas | BSS: ~0.89 mPas
Molecular Weight (MW)• Directly determines viscosity and tissue coating ability Higher MW → more cohesive | Lower MW → more dispersive• HA MW: 1–4 million Daltons typically
Osmolality• Must approximate aqueous humour (~305 mOsm/kg) Deviations cause corneal endothelial stress• All commercial OVDs: 270–310 mOsm/kg
pH• Should approximate physiological pH 7.0–7.4 Extreme pH → endothelial damage• Commercial OVDs: pH 6.8–7.8
Surface Tension• Lower surface tension → better spreading/coating Important for endothelial protection• Chondroitin sulfate has very low surface tension → excellent coater
Cohesion• Tendency of OVD molecules to stick to each other High cohesion → easier complete removal• Measured by retention/elution studies
Adhesion• Tendency to adhere to ocular surfaces/cells High adhesion → better endothelial protection• Dispersive OVDs have higher adhesion to tissues

Free preview ends here

Keep reading Ophthalmic Viscosurgical Devices (OVDs)

12 more sections are included with the Clinical plan.

  • Overview
  • OVD in Corneal Transplantation
  • Individual OVD Agents - Detailed Profiles
  • OVD in Glaucoma Surgery
  • OVD in Vitreoretinal Surgery
  • OVD in Paediatric Ophthalmology

Clinical plan

Every Learn topic in full, with oral-exam questions and MCQs.

View plans