OCT in Glaucoma: Structural Assessment

Glaucoma

Teaching images

11

1. Physics & Technology of OCT

CLINICAL ANALOGY: The Ultrasound of Light
  • Optical coherence tomography (OCT) is to the optic nerve what ultrasound B-scan is to the posterior segment - but instead of sound waves, Optical coherence tomography (OCT) uses near-infrared light (820-1310 nm).
  • Just as ultrasound measures echo time-delay of sound to reconstruct tissue depth, Optical coherence tomography (OCT) measures the echo time-delay of light using low-coherence interferometry.
  • The key advantage: light travels ~200,000x faster than sound, enabling axial resolution of 3-7 micrometers - fine enough to resolve individual retinal layers that no ultrasound could.

Core Physical Principle: Low-Coherence Interferometry

  • Optical coherence tomography (OCT) employs a Michelson interferometer with a broadband (low-coherence) near-infrared light source split into two arms: the reference arm (known path length to a mirror) and the sample arm (directed to the tissue).
  • Interference occurs only when the optical path difference is within the coherence length of the source.
  • In time-domain optical coherence tomography (OCT), scanning the reference-arm length reconstructs backscattered intensity at each tissue depth to generate an A-scan (depth profile).
  • Multiple A-scans constitute a B-scan (cross-sectional image).
Generation Comparison: TD-OCT vs SD-OCT vs SS-OCT
Generation Comparison: TD-OCT vs SD-OCT vs SS-OCT
ParameterTime-domain optical coherence tomography (TD-OCT) (1st Generation)Spectral-domain optical coherence tomography (SD-OCT) (2nd Generation)Swept-source optical coherence tomography (SS-OCT) (3rd Generation)
Mechanism• Moving reference mirror• Fourier-domain (spectrometer)• Fourier-domain (tunable laser)
Speed• Approximately 400 axial scans (A-scans)/sec• Approximately 20,000–100,000 axial scans (A-scans)/sec • Device-dependent• Approximately 100,000–400,000 axial scans (A-scans)/sec
Axial Resolution• Approximately 10 µm• 5–7 µm• 5–8 µm
Wavelength• 820 nm• 820–870 nm• 1050 nm
Sensitivity Roll-off• No Fourier-domain detector roll-off • Depth-related signal attenuation remains possible• Depth-dependent sensitivity roll-off• Generally reduced roll-off over greater imaging depths
Deep Tissue• Limited• Improved imaging, especially with enhanced-depth imaging• Generally improved penetration and visualization of the choroid and lamina cribrosa • Deep laminar imaging may still be attenuated
Example Device• Stratus OCT• Spectralis • Cirrus • RTVue• DRI Triton • PLEX Elite
VIVA PEARL
  • Spectral-domain optical coherence tomography (SD-OCT) is substantially faster than time-domain optical coherence tomography (TD-OCT), enabling dense volumetric scanning; device-specific eye tracking can reduce motion artifact.
  • Swept-source optical coherence tomography (SS-OCT) at 1050 nm penetrates deeper (choroid, lamina cribrosa) due to reduced scattering at longer wavelengths.
  • The sensitivity roll-off in spectral-domain optical coherence tomography (SD-OCT) means deep structures are imaged with lower sensitivity - a recognized limitation.
  • Optical coherence tomography (OCT) axial resolution is determined by light-source bandwidth and center wavelength; lateral resolution is determined by focal-spot/beam optics and is device-dependent.

Key Difference for Exams

  • Spectral-domain optical coherence tomography (SD-OCT): Spectrometer detects the interference spectrum simultaneously - fast but suffers sensitivity roll-off with depth.
  • Swept-source optical coherence tomography (SS-OCT): Tunable laser sweeps through wavelengths + single photodetector - better sensitivity at depth, superior for lamina cribrosa and choroidal imaging in glaucoma.
  • For glaucoma imaging of the lamina cribrosa, scleral canal, and other deep optic nerve head structures, swept-source optical coherence tomography (SS-OCT) can provide improved visualization because of reduced depth-dependent sensitivity loss and longer-wavelength imaging; it should not be described universally as the preferred modality.

2. OCT Anatomy: Structural Parameters

Exam pearl
  • CLINICAL ANALOGY: The Optic Nerve as a Fiber-Optic Cable Bundle
  • Imagine the optic nerve as a fiber-optic cable bundle.
  • The individual glass fibers represent retinal ganglion cell axons.
  • The supportive meshwork surrounding and separating the bundle is the scleral canal/lamina cribrosa.
  • In glaucoma, retinal ganglion cell axons are progressively damaged, particularly as they pass through the lamina cribrosa.
  • Optical coherence tomography (OCT) allows us to measure retinal nerve fiber layer thickness as a structural surrogate for axonal loss and assess optic nerve head morphology.
  • The thinner the retinal nerve fiber layer, the greater the structural loss of retinal ganglion cell axons.

Anatomical Architecture Imaged by OCT

  • Optical coherence tomography (OCT) in glaucoma targets three primary structural regions: the peripapillary retinal nerve fiber layer (pRNFL), the optic nerve head (ONH) including the lamina cribrosa, and the macular ganglion cell complex (mGCC).
  • Each provides complementary information about different stages of glaucomatous damage.
OCT ParameterAnatomical BasisGlaucoma RelevanceInstrument
Peripapillary retinal nerve fiber layer (RNFL) thickness• Reflects retinal ganglion cell (RGC) axonal tissue around the optic nerve head (ONH)• Useful for detecting structural glaucomatous damage • Should not be designated universally as the earliest detectable change• Spectral-domain optical coherence tomography (SD-OCT) • Swept-source optical coherence tomography (SS-OCT)
Rim area / neuroretinal rim (NRR)• Neural tissue within the disc margin• Structural rim narrowing or loss• Spectralis • Cirrus
Bruch's membrane opening–minimum rim width (BMO-MRW)• Neural tissue measured from the Bruch's membrane opening (BMO) to the internal limiting membrane• Geometry-aware neuroretinal rim measurement • Generally equal or better diagnostic capability than conventional two-dimensional rim area parameters• Spectralis HRA+OCT
Cup/disc ratio• Structural optic-disc cupping• Gross, disc-size-dependent assessment • Limited standalone sensitivity • Interpret with neuroretinal rim and other structural and functional findings• Optical coherence tomography (OCT) instruments
Macular ganglion cell complex (GCC) thickness• Macular retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), and inner plexiform layer (IPL)• Helps detect early and preperimetric glaucomatous damage • Complementary to peripapillary retinal nerve fiber layer (RNFL) assessment• Optovue RTVue
Ganglion cell-inner plexiform layer (GCIPL) / ganglion cell complex (GCC)• Macular ganglion cell layer (GCL), inner plexiform layer (IPL), and, for ganglion cell complex (GCC), macular retinal nerve fiber layer (RNFL)• Complementary to peripapillary retinal nerve fiber layer (RNFL) assessment• Cirrus provides ganglion cell-inner plexiform layer (GCIPL) and ganglion cell complex (GCC) measurements • Optovue RTVue provides ganglion cell complex (GCC) measurements
Lamina cribrosa• Collagenous connective-tissue plate at the optic nerve head (ONH)• Structural deformation, including posterior displacement, occurs in high-tension glaucoma (HTG) and normal-tension glaucoma (NTG)• Swept-source optical coherence tomography (SS-OCT) can assess lamina cribrosa morphology
Bruch's membrane opening (BMO) area• Size of the Bruch's membrane opening (BMO)• Contributes to the geometric characterization and size adjustment of Bruch's membrane opening–minimum rim width (BMO-MRW) and other rim measurements• Spectralis

The Retinal Layers - Nomenclature (International Nomenclature for Optical Coherence Tomography (IN•OCT) Consensus)

  • The International Nomenclature for Optical Coherence Tomography (IN•OCT) Panel (Staurenghi et al., 2014) proposed standardized nomenclature for retinal and choroidal layers and bands visible on spectral-domain optical coherence tomography images.
  • For glaucoma, the following layers are critical:

Glaucoma-Relevant Retinal Layer Stack (Inner to Outer)

  • Internal Limiting Membrane (ILM) - innermost boundary
  • Retinal Nerve Fiber Layer (RNFL) - retinal ganglion cell axons; THINNED in glaucoma
  • Ganglion Cell Layer (GCL) - retinal ganglion cell somata; LOST in glaucoma
  • Inner Plexiform Layer (IPL) - retinal ganglion cell dendrites + synapse layer
  • Inner Nuclear Layer (INL) - bipolar, amacrine, Müller cells
  • Outer Plexiform Layer (OPL)
  • Outer Nuclear Layer (ONL) - photoreceptor nuclei; preserved in glaucoma
  • External Limiting Membrane (ELM)
  • Inner Segment/Outer Segment junction (IS/OS junction; ellipsoid zone) - photoreceptor integrity
  • Retinal Pigment Epithelium (RPE)
VIVA PEARL
  • In glaucoma, the retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), and inner plexiform layer (IPL) are damaged.
  • The outer nuclear layer (ONL), ellipsoid zone (EZ), and retinal pigment epithelium (RPE) are generally spared in glaucoma.
  • The ganglion cell complex (GCC) comprises the retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), and inner plexiform layer (IPL); it is measured by RTVue/Optovue.
  • Cirrus ganglion cell–inner plexiform layer (GCIPL) comprises the ganglion cell layer (GCL) and inner plexiform layer (IPL). The ganglion cell complex (GCC) comprises the retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), and inner plexiform layer (IPL).
  • Macular retinal nerve fiber layer (RNFL) contribution: The inner ring of the macula contributes approximately 40% of retinal ganglion cell (RGC) axons to the optic nerve.

Retinal Nerve Fiber Layer (RNFL) Anatomy: The Inferior > Superior > Nasal > Temporal Rule & Arcuate Anatomy

Teaching image

1
  • The retinal nerve fiber layer (RNFL) is thickest in the superior and inferior peripapillary regions (double-hump pattern).
  • Arcuate fiber bundles from the temporal retina course above and below the macula to enter the superior and inferior regions of the optic disc.
  • The Inferior > Superior > Nasal > Temporal rule describes a commonly reported average retinal nerve fiber layer (RNFL) thickness pattern, although normal eyes do not uniformly follow it.
  • Glaucoma can violate this pattern through thinning in the inferior and/or superior arcuate regions; inferotemporal retinal nerve fiber layer (RNFL) defects may occur more frequently, but the first affected region varies between eyes.
QuadrantNormal Thickness (µm)First AffectedCorresponding VF Defect
Inferior (I)• ~125–130 µm • Device- and population-specific• Often affected early • Early damage may involve either the inferotemporal or superotemporal region• Superior arcuate / Bjerrum scotoma
Superior (S)• ~120–125 µm • Device- and population-specific• Often affected early • Early damage may involve either the inferotemporal or superotemporal region• Inferior arcuate scotoma
Nasal (N)• ~80 µm in some spectral-domain optical coherence tomography cohorts • Device- and population-specific• Late• Temporal visual-field loss (rare)
Temporal (T)• ~70 µm • Device- and population-specific• Usually less affected than the superior and inferior poles • Interpret temporal-sector involvement by phenotype and device-specific normative data• Central/paracentral
Clock-hour 7 (inf-temp)• ~140–145 µm in some spectral-domain optical coherence tomography cohorts • Device- and population-specific• Important inferotemporal sector • Often among the most discriminative clock-hour regions in glaucoma• Superior nasal step
Clock-hour 11 (sup-temp)• ~140 µm in some spectral-domain optical coherence tomography cohorts • Device- and population-specific• Important superotemporal sector • Among the clock-hour regions that can discriminate glaucoma• Inferior nasal step
Notes
  • • Retinal nerve fiber layer thickness values are approximate and vary with device, scan protocol, age, segmentation, and population. • Inferior and superior retinal nerve fiber layer thinning generally corresponds to superior and inferior visual-field loss, respectively; individual structure–function patterns vary.
Exam pearl
  • EXAM TRAP: The inferior retinal nerve fiber layer (RNFL) is thickest in normal eyes, so it has MORE tissue to lose before hitting the abnormal range.
  • This means: inferior damage can be extensive before crossing the red line on printouts.
  • Never dismiss superior visual field (VF) changes due to 'borderline' inferior optical coherence tomography (OCT) changes - look at absolute values and progression.

Free preview ends here

Keep reading OCT in Glaucoma: Structural Assessment

8 more sections are included with the Clinical plan.

  • RNFL Analysis: Interpretation Protocol
  • ONH & BMO-MRW Parameters
  • GCC / Macular Analysis in Glaucoma
  • OCT Angiography (OCTA) in Glaucoma
  • Glaucoma Progression Analysis on OCT
  • OCT Artifacts & Pitfalls

Clinical plan

Every Learn topic in full, with oral-exam questions and MCQs.

View plans