Glaucoma Drainage Device (Tube Shunt) Surgery
Glaucoma
Teaching images
101. Chapter 1
Teaching images
2Introduction, History & Clinical Analogy
Clinical Analogy: The Drainage Pipe System
- Clinical Analogy: Imagine the eye as a sink constantly receiving water (aqueous humor). In glaucoma, the drain (trabecular meshwork) is blocked. Trabeculectomy creates a new drain hole - but the hole may scar over. A tube shunt is like installing a permanent drainage pipe that bypasses the blocked drain entirely, leading water to a reservoir (bleb over the plate) where it absorbs safely. The tube = the pipe; the plate = the reservoir. The aqueous cannot scar the pipe shut the way it can block a new hole - hence, tube shunts work where trabeculectomy has failed.
Definition & Terminology
- Tube shunts - also called glaucoma drainage devices (GDDs), aqueous shunts, or setons - are implantable devices that create an alternate drainage pathway for aqueous humor from the anterior chamber (or posterior segment) to a subconjunctival reservoir, lowering intraocular pressure (IOP) in refractory glaucoma.
- The term 'seton' (from Latin, meaning 'bristle') reflects the historical use of silk threads as drainage conduits.
Exam Pearls
Oral-exam tip
- Molteno introduced the first clinically practical glaucoma drainage device (GDD) in 1969. The key innovation was a large surface area episcleral plate to disperse aqueous through a plate-associated bleb or capsule and provide a defined drainage surface - still a fundamental design principle of many conventional modern glaucoma drainage devices (GDDs).
- The terms 'glaucoma drainage device,' 'aqueous shunt,' and 'tube shunt' are used as overlapping terms in glaucoma literature; no universal preference for 'glaucoma drainage device' over the others is established.
2. Relevant Anatomy, Device Types & Components
Teaching images
2Relevant Surgical Anatomy
- Understanding tube shunt surgery requires mastery of the following anatomical regions:
- Anterior Chamber (AC) Angle: Tube tip ideally placed 1–2 millimetres (mm) beyond limbus into anterior chamber (AC); tip bevel UP to prevent iris occlusion; avoid corneal endothelium (minimum 2 millimetres (mm) clearance).
- Limbal Region (Corneoscleral Junction): Tube enters through a 23-gauge track made by needle. Limbus is 0.5–1 millimetres (mm) wide - entry too anterior → corneal damage; too posterior → ciliary body trauma.
- Episcleral Space (Sub-Tenon's): Plate seated 8–10 millimetres (mm) posterior to limbus between two rectus muscles; ideal position = superotemporal quadrant (broadest space, away from obliques).
- Tenon's Capsule: Fascial sheath of the eyeball extending from the optic nerve to the corneal limbus; fibrosis of the capsule around the plate can reduce permeability and contribute to treatment failure.
- Rectus Muscles: Medial rectus insertion 5.5 millimetres (mm), lateral 6.9 millimetres (mm), superior 7.7 millimetres (mm), inferior 6.5 millimetres (mm) from limbus - plate must be placed posterior to muscle insertions.
- Vortex Veins: Exit at equator 4 quadrants ~14–15 millimetres (mm) from limbus; avoid during plate dissection to prevent choroidal hemorrhage.
- Posterior Chamber Placement: Pars plana tube insertion through a 23-gauge tract 3.5 millimetres (mm) posterior to the limbus; thorough pars plana vitrectomy is performed before insertion in non-vitrectomized eyes to prevent postoperative tube occlusion by residual vitreous.
Components of a Tube Shunt
| Component | Description | Clinical Relevance |
|---|---|---|
| Tube (Silicone) | • Outer diameter 0.63 mm • Inner diameter 0.30 mm • Length trimmed at surgery | • Too long → tube–cornea/endothelial touch • Too short → iris occlusion and possible postoperative tube retraction from the anterior chamber |
| Episcleral Plate | • Material varies by device, including silicone, polymethylmethacrylate (PMMA), and polypropylene • Reported plate areas include 96–350 mm², with some models approximately 364 mm² | • Larger plate area may provide greater filtration surface and potential intraocular-pressure (IOP) lowering • Benefit is not consistently demonstrated • Larger or higher-profile devices may increase diplopia or motility complications |
| Valve Mechanism | • Present (Ahmed glaucoma valve [AGV]) • Absent (Molteno, Baerveldt) | • Valve prevents early hypotony • Non-valved needs ligation strategy |
| Fixation Eyelets | • Fixation eyelets on plate for suturing to the sclera | • Prevents plate migration • Typically 2 non-absorbable sutures |
| Patch Graft | • Pericardium, donor sclera, or cornea over tube | • Helps prevent tube erosion through conjunctiva |
| Component | Description | Clinical Relevance |
|---|---|---|
| Tube (Silicone) | • Outer diameter 0.63 mm • Inner diameter 0.30 mm • Length trimmed at surgery | • Too long → tube–cornea/endothelial touch • Too short → iris occlusion and possible postoperative tube retraction from the anterior chamber |
| Episcleral Plate | • Material varies by device, including silicone, polymethylmethacrylate (PMMA), and polypropylene • Reported plate areas include 96–350 mm², with some models approximately 364 mm² | • Larger plate area may provide greater filtration surface and potential intraocular-pressure (IOP) lowering • Benefit is not consistently demonstrated • Larger or higher-profile devices may increase diplopia or motility complications |
| Valve Mechanism | • Present (Ahmed glaucoma valve [AGV]) • Absent (Molteno, Baerveldt) | • Valve prevents early hypotony • Non-valved needs ligation strategy |
| Fixation Eyelets | • Fixation eyelets on plate for suturing to the sclera | • Prevents plate migration • Typically 2 non-absorbable sutures |
| Patch Graft | • Pericardium, donor sclera, or cornea over tube | • Helps prevent tube erosion through conjunctiva |
Major GDD Types & Comparison
| Parameter | Ahmed Glaucoma Valve (FP7/FP8) | Baerveldt (350/250) | Molteno (Single/Double) | Krupin |
|---|---|---|---|---|
| Plate Area | • 184 mm² (FP7) • 96 mm² (FP8) | • 350 mm² • 250 mm² | • 135 mm² (single) • 270 mm² (double) | • ~180 mm² |
| Valve | • Yes • Venturi mechanism • Silicone membrane | • No | • No | • Yes • Pressure-sensitive slit |
| Material | • Silicone | • Silicone plate | • Polypropylene | • Silicone |
| Intraocular Pressure Range | • Designed minimum intraocular pressure: ~8 mmHg • Opening and closing pressures vary | • Nonvalved • Temporary ligation limits early hypotony • Flow begins after ligature dissolution • Fenestration may provide controlled early flow | • Capsular resistance around the plate contributes to intraocular pressure control | • Designed to open: 11 mmHg • Designed to close: 9 mmHg • Opening and closing pressures may be inconsistent |
| Ligation Needed? | • No • Valve present | • Yes • Temporary tube ligation (e.g., 7-0 polyglactin [Vicryl]) • Proximal venting slits/fenestrations may supplement ligation | • Yes | • No |
| Final Mean Intraocular Pressure | • 14.7 mmHg at 5 years in the Ahmed Baerveldt Comparison Study | • 12.7 mmHg at 5 years in the Ahmed Baerveldt Comparison Study | • Outcomes vary by study and follow-up | • Outcomes vary by study and follow-up |
| Key Randomized Controlled Trial | • Ahmed Baerveldt Comparison Study (Ahmed vs Baerveldt) | • Tube Versus Trabeculectomy Study (vs trabeculectomy) | • - | • - |
| Preferred For | • High intraocular pressure • Immediate postoperative drainage desired | • Very refractory glaucoma • Lower final intraocular-pressure target | • Pediatric or neovascular glaucoma • Severe glaucoma uncontrolled by medical and conventional surgical treatment • Plate size/configuration individualized | • Valved implant • No uveitis-specific preference established |
| Parameter | Ahmed Glaucoma Valve (FP7/FP8) | Baerveldt (350/250) | Molteno (Single/Double) | Krupin |
|---|---|---|---|---|
| Plate Area | • 184 mm² (FP7) • 96 mm² (FP8) | • 350 mm² • 250 mm² | • 135 mm² (single) • 270 mm² (double) | • ~180 mm² |
| Valve | • Yes • Venturi mechanism • Silicone membrane | • No | • No | • Yes • Pressure-sensitive slit |
| Material | • Silicone | • Silicone plate | • Polypropylene | • Silicone |
| Intraocular Pressure Range | • Designed minimum intraocular pressure: ~8 mmHg • Opening and closing pressures vary | • Nonvalved • Temporary ligation limits early hypotony • Flow begins after ligature dissolution • Fenestration may provide controlled early flow | • Capsular resistance around the plate contributes to intraocular pressure control | • Designed to open: 11 mmHg • Designed to close: 9 mmHg • Opening and closing pressures may be inconsistent |
| Ligation Needed? | • No • Valve present | • Yes • Temporary tube ligation (e.g., 7-0 polyglactin [Vicryl]) • Proximal venting slits/fenestrations may supplement ligation | • Yes | • No |
| Final Mean Intraocular Pressure | • 14.7 mmHg at 5 years in the Ahmed Baerveldt Comparison Study | • 12.7 mmHg at 5 years in the Ahmed Baerveldt Comparison Study | • Outcomes vary by study and follow-up | • Outcomes vary by study and follow-up |
| Key Randomized Controlled Trial | • Ahmed Baerveldt Comparison Study (Ahmed vs Baerveldt) | • Tube Versus Trabeculectomy Study (vs trabeculectomy) | • - | • - |
| Preferred For | • High intraocular pressure • Immediate postoperative drainage desired | • Very refractory glaucoma • Lower final intraocular-pressure target | • Pediatric or neovascular glaucoma • Severe glaucoma uncontrolled by medical and conventional surgical treatment • Plate size/configuration individualized | • Valved implant • No uveitis-specific preference established |
Device Pearls - KEY DIFFERENTIATOR - Valved vs Non-Valved:
- Valved (Ahmed glaucoma valve (AGV)): Immediate intraocular pressure (IOP) control; risk = hypertensive phase (encapsulation of bleb at 4–8 weeks).
- Non-valved (Baerveldt, Molteno): Risk of early hypotony - needs tube ligation or fenestration.
- Non-valved implants may provide lower long-term intraocular pressure (IOP) than Ahmed, but implant selection should be individualized; evidence does not establish a universal below 14 millimeters of mercury threshold.
- FP7 = Adult (184 square millimeters), FP8 = Pediatric/Small eye (96 square millimeters) - memorize this!
- Baerveldt 350 has one of the largest plate areas among commonly used standard glaucoma drainage implants; 500-square-millimeter Baerveldt models have also been studied.
Free preview ends here
Keep reading Glaucoma Drainage Device (Tube Shunt) Surgery
9 more sections are included with the Clinical plan.