Interferometry and Optical Coherence

Optics & Refraction

Teaching images

3

1. Introduction - The Physics Beneath the Clinic

Exam pearl
  • Clinical Analogy: Think of two stone-drops in a still pond. Where the ripples overlap, they either amplify (constructive interference) or cancel (destructive interference). Light behaves identically - and ophthalmology has weaponised this principle into instruments that measure the eye with nanometre precision.
  • Interferometry is the science of using the interference of light waves to make ultra-precise measurements.
  • In ophthalmology, it underpins:
  • Optical Coherence Tomography (OCT) - cross-sectional retinal imaging
  • Laser Interferometry / Coherence Interferometry - preoperative visual potential assessment
  • Partial Coherence Interferometry (PCI) - biometry (IOLMaster, Lenstar)
  • Specular Microscopy - corneal endothelial imaging
  • Wavefront Sensing (Hartmann-Shack) - aberrometry

2. Fundamental Wave Physics

Wave Properties of Light

Teaching image

1
  • Light behaves as a transverse electromagnetic wave characterised by: wavelength (λ), frequency (f), amplitude (A), and phase (φ).
  • Two waves interact to produce interference.

Interference - Constructive vs Destructive

Teaching image

1
ParameterConstructive InterferenceDestructive Interference
Phase difference• 0° • 360° • 720° • (multiples of 2π)• 180° • 540° • (odd multiples of π)
Path differencenλ (n = 0,1,2,3…)(n + ½)λ
Resultant amplitudeA1 + A2 (maximum)|A1 − A2| (minimum/zero if equal)
Resultant intensityMaximum - bright fringeMinimum - dark fringe
Clinical useSignal peaks in optical coherence tomography (OCT)Null zones used in interferometry
Exam pearl
  • Intensity ∝ Amplitude². Equal amplitudes in destructive interference → complete cancellation. This is exploited in anti-reflection coatings on spectacle lenses.

Coherence - The Critical Concept

Teaching image

1
Exam pearl
  • Clinical Analogy: Coherence is like two singers staying perfectly in tune.
  • Temporal coherence = the electric field at one point maintains a stable phase relationship over time.
  • Spatial coherence = electric fields at different positions across the beam maintain a stable phase relationship.
  • A sufficiently coherent source, with spatial and temporal overlap and suitable polarization, produces stable, measurable interference fringes; reduced coherence lowers fringe visibility and may wash out the pattern.

Coherence has two complementary aspects:

  • Temporal Coherence: Phase relationship maintained over time → characterised by coherence length (Lc)
  • – Coherence length (Lc) ≈ λ²/∆λ (where ∆λ = spectral bandwidth; the numerical factor depends on the spectral shape and definition used)
  • – Narrow bandwidth (e.g., single-frequency laser) → long coherence length (Lc) → high temporal coherence
  • – Broad bandwidth (superluminescent diode (SLD)/white light) → short coherence length (Lc) → reduced temporal coherence - used in optical coherence tomography (OCT)!
  • Spatial Coherence: Phase relationship maintained across the wavefront width
  • – Point source → high spatial coherence
  • – Extended incoherent source → low spatial coherence
PropertyLaser (helium–neon, diode)Superluminescent diodeWhite light
Bandwidth• Narrow (source dependent) [14]• Moderate (20–100 nm) [14]• Very broad (source dependent) [14]
Coherence length• Long (source dependent) [15]• Short (source dependent) [15]• Very short (~1 µm)
Temporal coherence• Very high [15]• Low [15]• Very low [16]
Spatial coherence• Very high [16]• High• Low (source dependent) [16]
Ophthalmic use• Laser interferometry [17]• Optical coherence tomography (time-domain and spectral-domain) [18]• Specular microscopy
Axial resolution• Poor depth discrimination• High axial (depth) resolution in optical coherence tomography [19]• Poor depth discrimination
MCQ Trap
  • Optical coherence tomography (OCT) uses a SHORT coherence length source (broad bandwidth superluminescent diode (SLD)), NOT a laser.
  • The short coherence length is what gives optical coherence tomography (OCT) its axial resolution.
  • A laser's long coherence length would make depth discrimination impossible.

Free preview ends here

Keep reading Interferometry and Optical Coherence

12 more sections are included with the Clinical plan.

  • The Michelson Interferometer - Foundation Principle
  • Laser Interferometry - Potential Acuity Assessment
  • Partial Coherence Interferometry (PCI) - Optical Biometry
  • OCT - Interferometric Cross-Sectional Imaging
  • Interference in Clinical Optics - Everyday Applications
  • Wavefront Sensing and Aberrometry